57
BIN Measurement (Using Multiple Judgment Standards)
4.3
BIN Measurement (Using Multiple Judgment
Standards)
In BIN measurement, multiple sets (up to 10, BIN0 to BIN9) of judgment standards (consisting of
upper limit and lower limit values) are set. After each measurement, the instrument compares the
measured value to multiple judgment standards and generates judgment results accordingly. This
function provides a convenient way to group measurement targets into ranks.
The BIN number corresponding to the applicable judgment standard is displayed on the screen,
and a signal can be output from the EXT I/O connector.
Measured values that do not fall under any BIN are indicated as
OB
(“out of bins”). This function
can be used with both auto-range operation and fixed-range operation.
BIN0 BIN1 BIN2 BIN3 BIN4 BIN5 BIN6 BIN7 BIN8 BIN9
IN
IN
IN
IN
OB
OB
OB
Upper limit value
Lower limit value
BIN0 BIN1 BIN2 BIN3 BIN4 BIN5 BIN6 BIN7 BIN8 BIN9
IN
IN
IN
IN
OB
IN
IN
OB
OB
IN
IN
IN
IN
Upper limit value
Lower limit value
4
Judging Measured V
alues
www.
.com
1.800.561.8187
Содержание DM7275 Series