![Hioki DM7275 Series Скачать руководство пользователя страница 134](http://html1.mh-extra.com/html/hioki/dm7275-series/dm7275-series_instruction-manual_2137275134.webp)
133
Connections (Instrument and Control Device)
Internal circuit architecture
•Use ISO_COM as the common pin for both input signals and output signals.
•If a large current will flow to common wiring, branch the output signal common wiring and input signal
common wiring near the ISO_COM pin.
NPN setting
8
1
2
10
11
9
27
ISO_5 V
ISO_COM
ISO_COM
ERR
HI
TRIG, IN0
(予約)
2 kΩ
1 kΩ
10 Ω
EXT I/O MODE
selector
NPN
Do not connect an external power
supply to pin 8.
Internal isolated common
(Isolated from the instrument’s protective ground)
Internal isolated
power supply
Instrument
Programmable controller, etc.
Output
Common
Input
Common
(Reserved)
PNP setting
2 kΩ
1 kΩ
10 Ω
EXT I/O MODE
selector
PNP
8
11
9
27
ISO_5 V
ERR
TRIG, IN0
1
2
10
(予約)
HI
ISO_COM
ISO_COM
Do not connect an external power
supply to pin 8.
Internal isolated common
(Isolated from the instrument’s protective ground)
Internal isolated
power supply
Instrument
Programmable controller, etc.
Output
Common
Input
Common
(Reserved)
External Control (EXT
I/O)
11
www.
.com
1.800.561.8187
Содержание DM7275 Series