![Hioki DM7275 Series Скачать руководство пользователя страница 173](http://html1.mh-extra.com/html/hioki/dm7275-series/dm7275-series_instruction-manual_2137275173.webp)
174
Q&A (Frequently Asked Questions)
No.
Issue
Items to check
Possible causes → Solutions
See
4-5
EOM signal is not
output.
Measured
value
Not updated
See No. 3-2 above.
EOM signal logic
(The EOM signal turns ON once
measurement is completed.)
-
EOM signal
setting
Pulse
The pulse output time is short and
the controller cannot detect the
EOM signal.
→ Increase the pulse output time
of the EOM signal or set the output
setting to "hold".
Hold
The measurement time is short,
and the interval during which the
EOM signal is OFF cannot be
detected.
→ Change the EOM signal output
setting to “pulse”.
4-6
The Hi, IN and Lo
signals are not
output.
Comparator
judgment
results
Are not displayed
→ See No. 1-3 above.