![Hioki DM7275 Series Скачать руководство пользователя страница 176](http://html1.mh-extra.com/html/hioki/dm7275-series/dm7275-series_instruction-manual_2137275176.webp)
177
Error Displays
Display
Description
Solution
ERR:063
Error while reading the
USB memory.
An error occurred while
reading the USB flash drive.
The file may be damaged.
Recover the file or use a
different USB flash drive.
ERR:064
Error while reading the
configuration file.
An error occurred while
reading a setting file in the
USB flash drive.
The file may be damaged.
Recover the file or use a
different USB flash drive.
ERR:065
File not found.
A valid file was not found in the
USB flash drive.
Specify a proper file.
ERR:070
No space available.
There is no free space in the
USB flash drive.
Delete unnecessary files to
secure free space.
ERR:071
Error occurred saving
the file.
An error occurred while saving
data in the USB flash drive.
The file may be damaged.
Recover the file or use a
different USB flash drive.
ERR:076
Error occurred deleting
the file.
An error occurred while
deleting data in the USB flash
drive.
The file may be damaged.
Recover the file or use a
different USB flash drive.
ERR:077
Unable to rename the file
because another file with
the same name already
exists.
The file name cannot be
changed as there are files with
the same file name.
Specify a different file name.
ERR:078
Error occurred renaming
the file.
An error occurred while
changing a file name in the
USB flash drive.
The file may be damaged.
Recover the file or use a
different USB flash drive.
ERR:079
Error while reading the
USB memory.
An error occurred while
reading the USB flash drive.
The file may be damaged.
Recover the file or use a
different USB flash drive.
ERR:080
Unable to enter the
adjustment mode.
The mode cannot be changed
to adjustment mode.
The Adjustment screen is not
available for use by end-users.
ERR:090
ROM check sum error.
Check sum of the program
ROM does not match.
Malfunction of the instrument.
Request service.
ERR:091
RAM error.
The RAM failed.
Malfunction of the instrument.
Request service.
ERR:092
Memory access error.
Turn off the power and
restart after a while.
Communication with memory
failed.
Turn OFF the power and turn it
ON again after some time.
ERR:093
Memory test error.
Memory failure.
Malfunction of the instrument.
Request service.
ERR:094
Adjustment data error.
Adjustment data is not correct. Malfunction of the instrument.
Request service.
ERR:095
Backup data error.
Backup data is not correct.
Settings have been reset.
Reconfigure measurement
conditions and other settings.
ERR:096
Failed to detect line
frequency. Select line
frequency.
The power frequency has not
been detected.
Check the voltage and
frequency of the power supply.
ERR:098
"The clock is not set.
Reset? (15-01-01
00:00:00)"
Clock has not been set.
Replace the backup battery
and set a clock.
ERR:099
Failed to detect line
frequency; will be set to
50 Hz.
The power frequency has not
been detected. The frequency
will be set to 50 Hz.
Check the voltage and
frequency of the power supply.
Maintenance and Service
14
www.
.com
1.800.561.8187
Содержание DM7275 Series
Страница 207: ...13 09 www com information itm com 1 800 561 8187...