Configuring External Input and Output
TRIG signal operation while the input filter is set to
ON
(Chatter)
Response
time
During measurement
Hold the input signal until the response time has passed.
EOM signal output type
You can choose to either hold the EOM signal’s output until the next trigger is received or output
the set pulse.
(Measurement screen)
MENU
>
EXT I/O
1
2
Select the output type.
HOLD
Holds the EOM signal after
measurement is complete
(default setting).
PULSE
Outputs a pulse with the set
width after measurement is
complete.
3
(After selecting
PULSE
)
Set the pulse width and touch
ENTER
.
˄
Increases the value by 1.
˅
Decreases the value by 1.
CLR
Deletes all characters.
CANCEL
Cancels the setting and returns to
the previous screen.
Valid setting range: 1 ms to 100 ms
(Default setting: 5 ms)
Measurement
processing
External Control (EXT
I/O)
11
www.
.com
1.800.561.8187