140
Timing Chart
(1) When the trigger source is set to
EXTERNAL
and EOM output is set to
HOLD
Inputting the TRIG signal will cause the EOM signal to turn off and measurement to start. When
measurement completes, the EOM signal will turn on and remain on until the next TRIG signal is
input.
See “11 EOM signal output type” (p. 137).
TRIG
EOM
T3
T0
OFF
ON
HI, IN, LO, ERR
OB, BIN0 – BIN9
ON
T1
ON/OFF
T4
T5
T7
T8
T2
OFF
OFF
OFF
OFF
Contact
check
Measurement
processing
Delay
Measurement Calculation
•The TRIG signal is ignored while the EOM signal is off (i.e., while measurement is in progress).
•After changing settings, for example to switch ranges, allow processing time (100 ms) to elapse
before inputting the TRIG signal.
•The instrument will output the EOM signal as soon as the judgment result (HI, IN, LO, ERR, or
BIN) has been finalized. If the connected external device’s input circuitry is characterized by a
slow response, it may take time for the judgment result to be captured after the EOM signal is
detected as having turned on.
See “11 Measurement process (starting measurement from an external device and loading
(2) When the trigger source is set to
EXTERNAL
and EOM output is set to
PULSE
The EOM signal will turn on when measurement completes. Once the EOM output pulse width (T9)
has elapsed, the EOM signal will return to the off state. Inputting the TRIG signal while the EOM
signal is on will cause the EOM signal to turn off and measurement to start.
See “EOM signal output type” (p. 137).
TRIG
EOM
T3
T0
OFF
ON
HI, IN, LO, ERR
OB, BIN0 – BIN9
ON
OFF
OFF
OFF
ON/OFF
T4
T5
T7
T2
T9
T1
T8
Contact
check
Measurement
processing
Delay
Measurement Calculation
www.
.com
1.800.561.8187