PROBLEM
TROUBLESHOOTING
Damage to
Insulation on Probe
or Introducer
Do not use. Discard immediately.
Water is not
circulating through
tubing during
COOLIEF* SINERGY*
pre-cooling, ON and
post-cooling states
• Ensure the Tube Kit is correctly connected to
the Probe.
• Ensure the Tube Kit has been correctly placed
in the pumphead.
• Ensure the burette reservoir has been filled.
• Visually inspect the Tube Kit tubing and joints
for leaks and occlusions.
• Ensure that the float ball in the burette is
floating and not occluding the outflow of
water from the burette.
• Ensure the pump tubing (thick-walled tubing
that is coming directly out of the bottom port
of the burette) is placed in the pumphead.
Water is not
dripping into the
burette
Check to see if water is running down the wall of
the burette.
Float is stuck on
bottom port of the
burette
Close the pumphead lid. Gently shake the burette
to try and loosen the ball from the bottom of the
burette.
The lid of the
burette cannot be
removed
Inject sterile water through the port of the lid, rather
than removing the lid.
COOLIEF* Tube Kit
breaks, is leaking or
is occluded
Immediately discard the Tube Kit.
Customer Service and Product Return Information
If you have any problems with or questions about
HALYARD*
Equipment, contact
our technical support personnel:
Halyard Health
5405 Windward Parkway
Alpharetta, GA 30004 USA
E-mail: [email protected]
1-844-425-9273 (1-844-HALYARD)
Notes
In order to return products under limited warranty you must have a return
authorization number before shipping the products back to Halyard Health.
Limited Warranties
Halyard Health warrants that these products are free from defects in original
workmanship and materials. If these products prove to be defective in original
workmanship or original materials, Halyard Health, in its absolute and sole
discretion, will replace or repair any such product, less charges for transportation
and labor costs incidental to inspection, removal or restocking of product.
This limited warranty applies only to original factory delivered products that have
been used for their normal and intended uses. Halyard Health’s limited warranty
shall NOT apply to Halyard Health’s products which have been repaired, altered
or modified in any way and shall NOT apply to Halyard Health’s products which
have been improperly stored or improperly installed, operated or maintained
contrary to Halyard Health’s Instructions. The warranty period for
HALYARD*
RF
Probes and RF Generator Connector Cables is 90 days from the date of purchase,
unless otherwise stated.
Disclaimer and Exclusion of Other Warranties
There are no warranties of any kind, which extend beyond the description of the
warranties above. Halyard Health disclaims and excludes all warranties, whether
expressed or implied, of merchantability or fitness for a particular use of purpose.
Limitation of Liability for Damages
In any claim or lawsuit for damages arising from alleged breach of warranty,
breach of contract, negligence, product liability or any other legal or equitable
theory, the buyer specifically agrees that Halyard Health shall not be liable for
damages for loss of profits or claims of buyer’s customers for any such damages.
Halyard Health’s sole liability for damages shall be limited to the cost to buyer of
the specified goods sold by Halyard Health to buyer which give rise to the claim
for liability.
The buyer’s use of this product shall be deemed acceptance of the terms and
conditions of these limited warranties, exclusions, disclaimers and limitations of
liability for money damages.