GSI Group - Laser Division
JK300HP/JK300P/JK125P
Pre-installation and User Manual
Issue 2.0
Page 60 of 172
Part No: 1EA300E01
5.3 Disconnect External Water Supply
Disconnect External Water Supply
WARNING
ENSURE WATER SUPPLY IS ISOLATED BEFORE DRAINING COOLING WATER
SYSTEM.
1
Turn OFF external water supply.
2
Place drainage bowl underneath hose connectors.
3
Loosen the hose clips and remove the hoses.
4
Allow water to drain from system into drainage bowl.
5
Check system is drained of water; wipe up spillage using paper towels.
5.4 Drain Internal Cooling System
Drain Internal Water:
CAUTION
Ensure water is fully drained from internal cooling system. Ensure no residual
water is left in pipework or pumping chambers. Failure to do so may result in
severe damage to equipment during shipping.
1
Refer to Section 9.4.2 and drain the coolant system.
2
Place drainage bowl below cooling system bay.
3
Remove filter bowl and empty into drainage bowl.
4
Remove de-ioniser bowl and empty into drainage bowl.
5
Bleed coolant from the system.
6
Replace filter and de-ioniser bowls.
7
Remove spillage with paper towels.