7.6.5 Alarm 4n05 : Power Supply Over Temperature
Meaning:
The power supply temperature switch has been activated.
Possible causes:
1.
Configurable high temperature alarm level set too high.
2.
Incorrect cooler valve setting.
3.
Cooler valve stuck in the closed position.
4.
Factory supply cooling water turned off.
5.
Insufficient flow from factory cooling water supply.
6.
Faulty power supply.
7.6.6 Alarm 4n06 : Power Supply Contactor In Fault
Meaning:
The power supply contactor is in when it should be out. The power supply should be off.
Possible causes:
1.
Faulty power supply.
7.6.7 Alarm 4n07 : Power Supply Failed to initialise
Meaning:
The Interlock Module timed out waiting for power supply n to response to an initialisation
command, and report that it had entered the initialised state during the laser start up
sequence.
Possible causes:
1.
Incorrect laser type setting.
2.
Loose laser control network connections.
3.
Faulty power supply.
4.
Faulty Interlock module.
5.
Faulty Laser Control Card.
7.6.8 Alarm 4n08 : Power Supply Contactor In Timeout
Meaning:
The power supply contactor did not engage within the timeout period during the laser start
up sequence.
Possible causes:
1.
Faulty power supply.
7.6.9 Alarm 4n09 : Power Supply Failed To Simmer
Meaning:
The Interlock Module timed out waiting for power supply n to enter, and report that it had
entered the simmer state during the laser start up sequence.
Possible causes:
1.
Incorrect laser type setting.