Pre-installation and User Manual
Issue 2.0
Page 36 of 172
Part No: 1EA300E01
2.7.4 Safety Features
Two safety features of the JK300HP/JK300P/JK125P Series Laser which protect personnel
from exposure to Laser radiation are:
•
Remote Interlock Connector.
•
Workstation Interlock.
2.7.5 Remote Interlock Connector
The Remote Interlock Connector enables an external Remote Safety Interlock to be fitted to
the interlock circuit of the Laser. The Remote Interlock Connector is a requirement of the
two principal Laser radiation standards:
Worldwide
IEC 60825-1 (Europe EN 60825-1).
North
America
FDA
Regulation 21 CFR Chapter J 1040.10.
When the terminals of the RIC are open circuit, exposure to Laser radiation is prevented.
This is caused by either the RIC plug being removed or the terminals of a Safety Interlock
Switch connected to the plug being open circuit.
In these circumstances, the interlock chain is broken, causing the Arc-lamp power circuits to
be de-energised and the Shutter closed.
Restoring electrical continuity at the RIC will not restart the Laser. To restart, a normal Laser
start-up Sequence has to be followed.
This facility is equivalent to a remote emergency stop control. It must not be used for
planned shutdown of the Laser.