7.5.6 Alarm 3n06: Cooler Module Relay Overload Fault
Meaning:
Circuit breaker supplying the pump has tripped.
Possible causes:
1. Faulty
pump.
2. Pump
overheated.
3. Loose
connection.
4.
Faulty pump connections.
7.5.7 Alarm 3n07: Low temperature fault
Meaning:
The coolant temperature has fallen below a configurable threshold.
Possible causes:
1.
Incorrect cooler valve setting.
2.
Cooler valve stuck in the open position.
3.
Configurable threshold set too high.
4. Loose
connections.
5.
Faulty temperature sensor
6.
Faulty Cooler Module.
7.5.8 Alarm 3n08: High temperature fault
Meaning:
The coolant temperature has raised above a configurable threshold.
Possible causes:
1.
Incorrect cooler valve setting.
2.
Cooler valve stuck in the closed position.
3.
Factory supply cooling water turned off.
4.
Insufficient flow from factory cooling water supply.
5.
Configurable threshold set too low.
6. Loose
connections.
7.
Faulty temperature sensor.
8.
Faulty Cooler Module.
7.5.9 Alarm 3n09: Very High temperature fault
Meaning:
The coolant temperature has raised above a fixed software temperature threshold.
Possible causes:
1.
Configurable high temperature alarm level set too high.
2.
Incorrect cooler valve setting.
3.
Cooler valve stuck in the closed position.
4.
Factory supply cooling water turned off.
5.
Insufficient flow from factory cooling water supply.