23
3.
Once making a wrong selection, you can push the knob on the event again to give up the selection. Select
"EXIT" to exit the menu and consequently the selection will come into effect.
4.4.10 SD operate
Please refer to the chapter related to
SD Recall
.
4.5 Machine version
Select the "VERSION" item in the "SYSTEM MENU". In the popping up menu, you can learn the software version of
the monitor.
Software name
CMS8000
CMS9000
(
CMS7000
)
Specification
None.
None.
Version
2.50311162128.66817
1.50311162128.66817
Naming standard
"Major adaptive upgrade", "Major enhancive software upgrade", "Major improvement
software upgrade", "Minor corrective software upgrades", "Build"
4.6 Drug calculation
You can use the drug calculation and titration table function of the monitor to calculate the concentration of 15 kinds of
drugs. Refer to the
Chapter Drug Calculation and Titration Table
for detailed information.
4.7 Maintain
4.7.1 User maintain
1. You need to select the "MAINTAIN" item in the "SYSTEM MENU", then select "USER KEY".
2. Input the password to enter the user maintain menu, then you can customize the maintenance settings. Items
shown as below can be set:
n
LANGUAGE: select the language you need
n
LEAD NAMING: AHA or EURO
n
HELP SETUP: ON/OFF
n
NIBP OBSTRUCT SETUP: 1/2/3/4
This function is used to detect whether the patient moves during the blood pressure measurement. If the
patient moves, the monitor will give an alarm message and stop the current measuring, or the measurement
will be taken as usual.
1)
This function is set "1" as default.
2)
"1" represents the sensitivity is reduced to the minimum, "4" represents the sensitivity is increased
to the maximum. The higher level the sensitivity is set and the easier to detect the interference of
movement.
n
"NETWORK CONFIGURATION": see
Section 4.7.3 Network Configuration
for details.
n
HL7 server configuration:
①
IP: 202.114.4.120. Input the IP address of the server.
②
Port: 511. Input the server port.
③
Sending interval: 1. Set the frequency of data sending, unit is “second”.
n
ALARM SET PASSWORD MODIFICATION: it is used to change the login password of "ALARM SETUP".
Содержание 35152