"NIBP CUFF TYPE
ERR"
Perhaps the used cuff does not fit
the patient type sey.
LOW
Check if the patient type is set up
correctly. Check whether the cuff
conforms to the patient's type.
"NIBP PNEUMATIC
LEAK"
NIBP airway has leaks.
LOW
Check the connection of each part or
replace with a new cuff. If the failure
still exists, contact the manufacturer for
repair.
"NIBP SYSTEM
FAILURE"
Problem happens when measuring
the curve. The system cannot
perform measurement, analysis or
calculation.
LOW
Check the connection of each part and
the patient situation. Measure again, if
the failure still exists, contact the
manufacturer for repair.