13.6 Amend results
Keep the limb to be measured and the patient's heart on one horizontal position. Otherwise amend the measurement
results by the following methods:
n
If the cuff is higher than the horizontal position of the heart, then the value should add 0.75 mmHg(0.10 kPa) after
the displayed value.
n
If the cuff is lower than the horizontal position of the heart, then the value should subtract 0.75 mmHg(0.10 kPa)
after the displayed value.
13.7 NIBP display
There is no waveform for NIBP measurement, it only displays the NIBP measurement results. The following figure is
only used for reference, your device may display a different interface.
1. Alarm is off
2. Measurement time
3. Mean pressure
4. Unit: mmHg or kPa
5. Diastolic pressure
6. Current cuff presure
7. Prompt information area: display the prompt information related to the NIBP.
8. Measurement mode
9. Systolic pressure
13.8 NIBP SETUP
Move the cursor to the NIBP hot key, press it to enter the "NIBP SETUP" interface.
n
ALM REC: select "ON" to enable report printing upon NIBP alarm.
n
Unit: mmHg or kPa
n
INTERVAL
Interval time in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes. After selecting the interval time, the
information "Please start" will appear in the NIBP prompt area, then press "NIBP" button to start the first auto
measurement. Select "MANUAL" in interval time to stop auto measuring and enter to manual measurement.
n
INFLATION
Press this button to select the initial pressure value for the cuff next time, there are different pre-inflation value ranges in
3
4
2
1
6
8
5
MANUAL
NIBP OVER PRESSURE
9
CUFF:100