![Gima 35152 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/gima/35152/35152_user-manual_2228007046.webp)
38
n
Press the switch to open the recorder door.
n
Insert a new roll of paper into the paper cassette, put the paper correctly and pay attention to the edges.
n
Give out the paper from the recorder outlet.
n
Close the recorder door.
NOTE:
l
Be careful when inserting paper. Avoid damaging the thermosensitive printer head. Unless replacing the
recorder paper or troubleshooting, do not leave the recorder door open.
Removing Paper Jam
When the recorder functions or sounds improperly, open the recorder door to check whether paper jam exists. If yes,
re-install the recorder paper.
Recorder Status Message (Technical Alarm)
Message
Cause
Alarm Level
Solution
RECORDER OUT OF
PAPER
Record paper runs out.
Low
Insert a new roll of record paper.
RECORDER ERROR
The communication of
recorder is abnormal.
Low
Tun off the monitor and restart it.
After restarting the monitor, if error still exists, contact our service engineers please.