91
ID1 LOW
DIA measuring value of channel 1 is below lower alarm limit.
User-selectable
IM1 HI
MAP measuring value of channel 1 is above upper alarm limit.
User-selectable
IM1 LOW
MAP measuring value of channel 1 is below lower alarm limit.
User-selectable
IS2 HI
SYS measuring value of channel 2 is above upper alarm limit.
User-selectable
IS2 LOW
SYS measuring value of channel 2 is below lower alarm limit.
User-selectable
ID2 HI
DIA measuring value of channel 2 is above upper alarm limit.
User-selectable
ID2 LOW
DIA measuring value of channel 2 is below lower alarm limit.
User-selectable
IM2 HI
MAP measuring value of channel 2 is above upper alarm limit.
User-selectable
IM2 LOW
MAP measuring value of channel 2 is below lower alarm limit.
User-selectable
Technical alarms:
Message
Cause
Alarm
Level
Remedy
IBP1 SENSOR OFF
IBP cable of channel 1
falls off from monitor.
LOW
Make sure that the cable is properly
connected.
IBP2 SENSOR OFF
IBP cable of channel 2
falls off from monitor.
LOW
Make sure that the cable is properly
connected.
IBP(1,2) INIT ERR
IBP module failure
HIGH
Stop using the measurement function
of IBP module, notify biomedical
engineer or Our service staff.
IBP(1,2) INIT ERR1
IBP(1,2) INIT ERR2
IBP(1,2) INIT ERR3
IBP(1,2) INIT ERR4
IBP(1,2) INIT ERR5
IBP(1,2) INIT ERR6
IBP(1,2) INIT ERR7
IBP(1,2) INIT ERR8
IBP(1,2) COMM STOP
IBP module failure or
communication failure
HIGH
Stop using the measurement function
of IBP module, notify biomedical
engineer or Our service staff.
IBP(1,2) COMM ERR
IBP module failure or
communication failure
HIGH
Stop using the measurement function
of IBP module, notify biomedical
engineer or Our service staff.
IBP1 ALM LMT ERR
Functional safety failure
HIGH
Stop using the measurement function
of IBP module, notify biomedical
engineer or Our service staff.
IBP2 ALM LMT ERR
Functional safety failure
HIGH
Stop using the measurement function
of IBP module, notify biomedical
engineer or Our service staff.