82
PRESSURE
specified upper safety limit.
measuring function of NIBP module and
notify biomedical engineer or Our service
staff.
NIBP SIGNAL
SATURATED
Excessive motion
LOW
Stop the patient from moving.
NIBP PNEUMATIC
LEAK
During pneumatic test, leak
is detected.
LOW
Check and replace the leaking parts, if
required, notify biomedical engineer or Our
service staff.
NIBP SYSTEM
FAILURE
Operation of blood pressure
pump system is failed.
HIGH
Stop using measuring function of NIBP
module, notify biomedical engineer or Our
service staff.
NIBP CUFF TYPE
ERROR
Cuff type does not comply
with the patient type.
LOW
Select an appropriate cuff type
NIBP TIME OUT
Measurement time has
exceeded 120 seconds
(adult) or 90 seconds
(neonatal).
HIGH
Measure again or use other measurement
methods.
NIBP ILLEGALLY
RESET
Abnormal module reset
HIGH
Reset again.
MEASURE FAIL
The system cannot perform
measurement, analysis or
calculation during
measuring.
HIGH
Check the cuff. Make sure that the patient
under monitoring is motionless. Measure
again.
Prompt message: (display in the prompt area below NIBP value)
Message
Cause
Alarm Level
Manual measure…
During manual measuring mode.
No alarm
Cont measuring…
During continuous measuring mode.
Auto measuring…
During automatic measuring mode.
Please start
After selecting interval time in MENU
Measurement over
Press NIBP key during measuring to stop measuring.
Calibrating…
During calibrating
Calibration over
Calibration over
Pneum testing…
During pneumatic test
Pneum test over
pneumatic test over
Resetting...
NIBP module in resetting
Reset failed
NIBP module reset failed