76
4.
Connect the cuff to the airway tube. Make sure that the airway tube is neither blocked nor tangled.
5.
Select a measurement mode in "NIBP SETUP" interface. Refer to the following paragraphs "Operation Hints" for
Details
6.
Press "NIBP" button on the front panel to start a measurement.
NOTE:
l
When measuring NIBP, keep subject and subject’s limbs still, not move and talk.
13.5 Operation hints
1.
Manual operation
n
Select "MANUAL" in "INTERVAL" item of "NIBP SETUP" interface, then press "NIBP" button on the front
panel to start a manual measurement.
n
During the idle time of auto measuring process, press "NIBP" button on the front panel to start a manual
measurement. Press "NIBP" button again to stop manual measurement and the system continues auto measuring.
2.
Auto measuring
Select a interval value in "INTERVAL" item of "NIBP SETUP" interface to perform auto measurement., then press
"NIBP" button on the front panel to start the first measurement, after finishing, the system will automatically measure
according to the interval time.
3.
Continuous measuring
Select "CONTINUAL" item in "NIBP SETUP" interface to start a continuous measurement. The precess will continue 5
minutes.
4.
Stop measuring
During measuring, press "NIBP" button on the front panel to stop measuring.
WARNING
l
In auto or continuous mode, if the time is too long, then the limb rubbed with the cuff may appear purpura,
ischemia and nerve injury. So when monitoring the patient, patient's limb color, warmth and sensitivity
should be checked frequently. Once any abnormality appears, please replace the cuff location or stop the
NIBP measurement.