Chapter 18 Maintenance and Cleaning
Only use the material and method listed in this chapter to clean or maintain the device. Otherwise we do not provide
any guarantee.
Our company has verified the cleaning and disinfection methods described in the manual. Professional personnel in
hospital should obey the manual to ensure sufficient cleaning and disinfection.
18.1 Introduction
Keep the device and accessories out of dust. In order to prevent damage, please obey the following rules:
n
Please dilute the detergent and disinfectant according to the manufacturer's instructions, or adopt the lower
concentration as soon as possible.
n
Don't immerse the device into the liquid.
n
Don't pour the liquid into the device or accessories.
n
Don't allow liquid to enter into the enclosures.
n
Don't use abrasion material(such as steel wool or silver polishing agent) and any strong solvent(such as acetone or
the detergent contained acetone).
18.2 Cleaning
The device should be cleaned periodically, in the area of seriously polluted or greater sand wind, cleaning frequency
should be increased. Before cleaning, please consult or understand the regulations about device cleaning in advance.
Selectable detergents:
n
Diluted ammonia water
n
Diluted sodium hypochlorite(Bleaching powder).
n
Hydrogen Peroxide(3%)
n
Alcohol(70%)
n
Isopropanol(70%)
When cleaning the device with the adsorption detergent, or wipe the residual detergent after cleaning, please use the
clean and non-corrosive soft cloth or paper towel.
18.2.1 Cleaning for host
Clean the device surface according to the following steps:
1
.
Turn off the power and unplug the power cord.
2
.
Use the soft cloth adsorbed proper detergent to completely wipe the external surface(including the LCD) of the
device until that there is no obvious dirt.
3
.
After cleaning, please use the new cloth or paper towel adsorbed proper tap-water to wipe the residual detergent
until that there is no obvious dirt.
4
.
Place the device in ventilation and shady environment for air drying.
WARNING
l
Before cleaning, make sure that the device is switched off and disconnected from the power cord.