package is intact before use, especially the disposable accessories. If any damage is found, please don't put is
into use.
NOTE:
l
Keep the package and packing materials for possible future transportation or storage.
3.2 Environmental requirement
Please obey the following instructions to ensure the safety of electrical installation. The environment for potable
monitor using shall properly away from vibration, dust, corrosive or flammable gas, extreme temperature or humidity
and so on. When it is installed in a cabinet, there should be enough space in front of the device for convenient operation.
When the door of the cabinet is opening, enough space at the back of the device should be guaranteed for convenient
maintenance. Allow at least 2 inches (5 cm) of space around the instrument to ensure air circulation.
WARNING
l
The environment for use, storage and transport should meets the requirements described in this manual,
otherwise the specifications of this product stated in this manual may not be able to achieved, or even cause
damage to the device.
Make sure that the device is free from condensation during working, when it is carried from one room to another room,
condensation may appears. This is because it is exposed under humid air with different temperatures.
3.3 Install the Monitor
If everything goes well, please place the monitor on a flat surface or fix it on the wall. The installation of wall bracket
please refer to its instructions.
3.3.1 Place on a Flat Surface
Place the monitor on a flat surface. The surface should be away from vibration, dust or corrosive drugs.
3.4 Connect the Power Cables
Please use the power cord equipped for the monitor. Plug the power cord to the power port on the monitor, and another
end to a grounded three-core power socket.
If the monitor is equipped with an adopter, plug one end of the adopter to the power port on the monitor, and another
end to a grounded three-core power socket.
NOTE:
l
Plug the power cord to the hospital outlet. If necessary, connect it with the equipotential ground wire.
l
When the device is equipped with battery, it must be charged after transport or storage. If turn on the
device directly without connecting with AC power supply, it may not work normally due to lack of electricity.
The device can be charged after connecting with the AC power no matter it is turned on or not.
Ground
In order to protect patients and medical personnel, the enclosure of portable monitor must be grounded. Therefore, the