73
measurement.
Physiological alarm:
Message
Cause
Alarm Level
SpO
2
HI
SpO
2
measurement value is higher than the upper limit of alarm.
User-selectable
SpO
2
LOW
SpO
2
measurement value is lower than the lower limit of alarm.
User-selectable
PR HI
PR measurement value is higher than the upper limit of alarm.
User-selectable
PR LOW
PR measurement value is lower than the lower limit of alarm.
User-selectable
Technical alarms:
Message
Cause
Alarm Level
Remedy
SpO
2
SENSOR
OFF
SpO
2
sensor may be
disconnected from the
patient or the monitor.
LOW
Make sure the sensor is placed in patient's
finger or other parts, and the connection
between the monitor and the cables is well.
SpO
2
COMM
ERR
SpO
2
module failure or
communication error.
HIGH
Stop using the measuring function of SpO
2
module, notify biomedical engineer or our
service staff.
SpO
2
SENSOR
FAULT
Inserting the probe falsely
will cause the short circuit
for SpO2 circuit or SpO2
probe cable.
HIGH
Check the type of SpO2 probe or replace the
SpO2 probe.
Prompt message:
Message
Cause
Alarm Level
SpO
2
SEARCHING PR
SpO
2
module is searching for pulse.
No alarm
SpO
2
SEARCH TIMEOUT
SpO
2
module cannot detect SpO
2
signal for a long
time.
HIGH