105
18.2.2.4 Cleaning for TEMP probe
1.
Use the soft cloth adsorbed proper detergent to wipe the contact position between the probe and the patient
until that there is no obvious dirt.
2.
After cleaning, please use the new cloth or paper towel adsorbed proper tap-water to wipe the residual
detergent until that there is no obvious dirt.
3. Use a dry soft to wipe the residual water.
4. Place the probe in ventilation and shady environment for air drying.
18.2.2.5 Cleaning for IBP cable
1.
Use the soft cloth adsorbed proper detergent to completely wipe the lead cable surface until that there is no
obvious dirt.
2.
After cleaning, please use the new cloth or paper towel adsorbed proper tap-water to wipe the residual
detergent until that there is no obvious dirt.
3. Use a dry soft to wipe the residual water.
4. Place the cable in a ventilation and shady environment for air drying.
18.3 Disinfection
To avoid extended damage to the device, disinfection is only recommended when stipulated as necessary in the Hospital
Maintenance Schedule. The device should be cleaned firstly before disinfection.
Disinfectant recommended: alcohol(70%), isopropanol(70%).
If alcohol or isopropanol is used in both cleaning and disinfection, please cleaning cloth for disinfection.