![Gima 35152 Скачать руководство пользователя страница 135](http://html1.mh-extra.com/html/gima/35152/35152_user-manual_2228007135.webp)
127
Measurement impedance
range
0.3~5
Ω
Base line impedance range
500~2500
Ω
Differential input impedance
>2.5 M
Ω
Bandwidth
0.2~2.5 Hz
Scan speed
6.25 mm/s, 12.5 mm/s, 25 mm/s
RR
Measurement range
0~150 rpm
Resolution
1 rpm
Accuracy
±2 rpm
Apnea alarm
10~40 s
Alarm limit
range(rpm)
Step(rpm)
Alarm high limit
(low limit+1)~150
1
Alarm low limit
Adult: 0~(high limit-1)
Pediatric and neonate: 0~(high limit-1)
A.11 NIBP
Measurement method
Oscillometric
Working mode
Manual, Auto, STAT
Measurement
Interval
in
AUTO Mode
1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 240, 480, 960 min
Measuring Period in STAT
Mode
5 min
Measurement parameters
SYS, DIA, MEAN
NIBP measurement
range(mmHg)
Adult
Pediatric
Neonatal
Systolic pressure
40~270
40~200
40~135
Diastolic pressure
10~215
10~150
10~100
Mean pressure
20~235
20~165
20~110
Accuracy
Maximum mean error: ±5 mmHg
Maximum Standard deviation: 8 mmHg
Pressure resolution
1 mmHg
Cuff pressure accuracy
±3 mmHg
Over-pressure protection
Adult: 297 mmHg
±
3 mmHg
Pediatric: 240 mmHg
±
3 mmHg
Neonate: 147 mmHg
±
3 mmHg
Alarm limit
Range(mmHg)
Step(mmHg)
High limit of systolic
pressure
Adult: (low limit+1)~270
Pediatric: (low limit+1)~200
1
Содержание 35152
Страница 12: ...4 CMS8000 CMS9000 1 11 9 10 2 6 12 3 4 5 7 8 12 10 4 9 7 1 11 8 6 5 3 2...
Страница 14: ...6 Side view CMS7000 CMS8000 CMS9000 1 2 3 5 6 7 8 4 9 10 1 2 3 5 6 7 8 4 9 1 0 1 2 3 5 6 7 8 4 9 10...
Страница 149: ...141 The table below shows ARMS values measured using SpO2 sensor S5RCS300 with Patient Monitor in a clinical study...