137
"T1 SENSOR OFF"
T1 sensor is not connected
correctly.
LOW
Check the connection of T1 sensor.
"T2 SENSOR OFF"
T2 sensor is not connected
correctly.
LOW
Check the connection of T2 sensor.
"ECG NOISE"
larger interference signals appear
in the ECG signals.
LOW
Check the connection of ECG lead
cable; Check the current situation of the
patient. Check if the patient moves a lot.
"XX COMM ERR"
XX module can't communicate
normally with the host.
HIGH
Re-start up the monitor, if the error still
exists, contact the manufacturer.
XX represents all the parameter modules in the system such as ECG, NIBP, SpO
2
, etc.
"KEYBOARD
COMM ERR"
The keyboard has failures, which
cannot be used.
HIGH
Contact the manufacturer for repair.
"WIRELESS
CONNECT ERROR"
The network part in the system
appear problems, so the system
can't achieve network function.
MED
Contact the manufacturer for repair.
"CAN'T
FIND
WIRELESS
DEVICE"
"LOW POWER"
Low battery.
HIGH
Charge the battery by connecting the AC
cord.
"RECORDER
OUT
OF PAPER"
No recording paper
LOW
Install the recording paper.
"RECORDER
ERROR"
The
recorder
communicates
abnormally.
LOW
Turn off the device and restart it.
"NIBP INIT ERR"
NIBP initialization error
LOW
Execute the reset program in the NIBP
menu. If the failure still exists, contact
the manufacturer for repair.
"NIBP
SELFTEST
ERR"
"NIBP ILLEGALLY
RESET"
During NIBP measurement, illegal
reset occurs.
LOW
Check the airway of NIBP to check if
there are clogs. Then measure again, if
the failure still exists, contact the
manufacturer for repair.