34
Chapter 7 Recording
NOTE:
l
The recorder is an optional component.
7.1 General Information for Recorder
A thermal array recorder is used for the Monitor.
Performance of the Recorder
n
Recording speed: 25 mm/s or 50 mm/s.
n
Waveform recording width: 48mm
n
It can record up to 2 waveforms.
n
The time and waveform of real-time recording are user-configurable.
n
Auto recording interval is set by user, the waveform is in accordance with the real time recording.
n
The alarm recording waveform is automatically selected by the monitor.
NOTE:
l
It is recommended to stop the recording when low battery alarm generated. Otherwise, the device may
shutdown for out of power.
7.2 Recording Type
The monitor provides several stripe recording types:
n
Continuous real-time recording
n
8 seconds real-time recording
n
Auto 8 seconds recording
n
Alarm recording
n
Freeze waveform recording
n
Trend graph/table recording
n
ARR review recording
n
Alarm recall recording
n
NIBP recall recording
n
SD recall recording
n
Drug calculation titration recording
Real-time Recording
Real-time recording starts as you pressing the "REC/STOP" button on the recorder.
The waveforms for continuous real-time recording and continuous 8 seconds recording are set in system setup (usually
the first two waveforms are displayed on the screen). You can also configure it through the menu. Refer to related
section for details.
In RECORD SETUP menu, user can choose to print two different waveforms at the same time, or print only one
waveform by setting the other waveform off. If two waveforms are set off, the real time record will print out measured
parameters only.