66
PVC
Without
pacemaker
Single PVC not belonging to the type of above mentioned
PVCs.
User-selectable
TACHY
All patients
5 consecutive QRS complex , RR interval is less than
0.5s.
User-selectable
BRADY
All patients
5 consecutive QRS complex, RR interval is longer than
1.5s.
User-selectable
MISSED
BEATS
Without
pacemaker
When HR is less than 100 beats/min, no heart beat is
tested during the period 1.75 times of the average RR
interval; or When HR is larger than 100 beats/min, no beat
is tested with 1 second.
User-selectable
PNP
With
pacemaker
No QRS complex and pacing pulse are available during
the period 1.75 times of the average R-R interval (only
considering patients with pacemaker.)
User-selectable
PNC
With
pacemaker
When pacing pulse is available, no QRS complex exists
during the period 1.75 times of the average RR interval
(only considering patients with pacemaker.)
User-selectable
Applicable patient type: "All patients" refers to perform Arr.analysis on patients either with pacemakers or without
pacemakers.
"Without pacemaker": refers to perform Arr. Analysis only on the patients without pacemakers.
"With pacemaker": refers to perform Arr. Analysis only on the patients with pacemakers.
Prompt message:
Message
Cause
Alarm Level
ARR LEARNING
The QRS template building required for Arr. Analysis is in process.
No alarm
NOTE:
l
Arrhythmia name displays in the alarm area.