Chapter 3 Installation
The portable monitor is designed to comply with relevant safety requirements of IEC 60601-1, IEC 60601-2-27 and
IEC 80601-2-30 for medical electrical equipment. The system has a floating input for defibrillation proof and
electrosurgical knife protection. If the correct electrodes (see the section about ECG Monitoring) are used and placed
according to the manufacturer's instructions, the display will be restored within 5 seconds after defibrillation.
WARNING
l
If any sign of damage to the monitor function is detected, or an error message appears, do not use it on any
patient. Contact biomedical engineer in the hospital or our maintenance engineer immediately.
l
All analog and digital equipment connected to this device must be certified by specified IEC standards (e.g.
IEC 60950 and IEC 60601-1), and all equipment shall comply with the requirements of IEC 60601-1-1 (valid
versions) for connection. The person who connects the additional equipment to the input/output port, is
responsible for the compliance with the IEC 60601-1-1 standard. If you have any questions, please contact
us.
l
When this device is connected to other electrical equipment in order to achieve a specific function, if the
hazards of this combination can not be determined from the specifications of each equipment (for example,
the risk of electric shock due to the accumulation of leakage current), please contact our company or
experts in the hospital related this field to ensure that the necessary safety of all equipment in this
combination will not be damaged.
l
Please use our designated bracket (optional). When installing the bracket, please avoid the screws to touch
the circuit board inside the machine.
NOTE:
l
To ensure the monitor works normally, please read this chapter and the content about patient safety before
use, and follow the requirements for installation.
l
If the monitor finds any fatal error during self-test, it will alarm.
l
Keep the package and packing materials for possible future transportation or storage.
3.1 Open the Package and Check
Before opening the package, please check it carefully. If any damage is found, please contact the carrier immediately.
Open the package and take out the monitor and accessories carefully. Check the components according to the packing
list to see whether the device has any mechanical damage or any part is missing. If there is any problem, contact the our
company immediately.
WARNING
l
The disposal of packaging materials should obey the local regulations or the hospital waste disposal system.
The packaging material must be keep out of the reach of children.
l
The device may get biological contaminated during storage, transport or use. Please confirm that the