Cylinder: one
Procedure of the air leakage test:
1. Set the "Patient type" to "Adult".
2.
Connect the cuff with the NIBP cuff jack.
3. Wrap the cuff around the cylinder of an appropriate size.
Diagram of NIBP Air Leakage Test
4. Select "PNEUMATIC" in NIBP menu, then the information "Pneum testing..." will display in the NIBP
parameter area.
5. The system will automatically inflate to 180 mmHg.
6. The system will automatically deflate after about 20s, it indicates that the air leakage test has finished.
7. If no prompt information appears in NIBP parameter area, it indicates that the airway is in good
situation and no air leaks exist. However if the prompt information “NIBP PNEUMATIC LEAK”
appears, it indicates that the airway may have air leaks. In this case, the user should check whether the
connection is loose. After confirming properconnections, the user should re-perform the pneumatic test.
If the failure prompt still appears, please contact the manufacturer for maintenance.
WARNING
l
This pneumatic test other than being specified in the EN 1060-1 standard is to be used by the user to simply
determine whether there are air leaks in the NIBP airway. If at the end of the test, the system prompts that
the NIBP airway has air leaks, please contact the manufacturer for maintenance.
n
DEFAULT: Select "DEFAULT" to enter "NIBP DEFAULT CONFIG" interface, the user may select "FACTORY
DEFAULT CONFIG" or "USER DEFAULT CONFIG". After selecting, the system will prompt for your
confirmation.
13.9 NIBP Calibration
When the user can not calibrate the NIBP, please contact with the maintenance personnel. The cuff pressure transducer
should be maintained by the qualified service personnel, and it should be checked and calibrated once per two years at
least.
13.10 NIBP Alarm Message
Physiological alarm belongs to the alarm which triggers by the parameters exceeding the limits, which may activate the
recorder to automatically output the parameters and related measured waveforms when the alarms occur on the
condition that the alarm record switch in the related menu is On.
Monitor
NIBP cuff jack
Airway tube
Cuff
Cylinder
Содержание 35152
Страница 12: ...4 CMS8000 CMS9000 1 11 9 10 2 6 12 3 4 5 7 8 12 10 4 9 7 1 11 8 6 5 3 2...
Страница 14: ...6 Side view CMS7000 CMS8000 CMS9000 1 2 3 5 6 7 8 4 9 10 1 2 3 5 6 7 8 4 9 1 0 1 2 3 5 6 7 8 4 9 10...
Страница 149: ...141 The table below shows ARMS values measured using SpO2 sensor S5RCS300 with Patient Monitor in a clinical study...