Description
43
The <Esc> key cancels unsaved changes to the monitor settings or
moves back to the next-higher menu.
4.4
Color LEDs (Light Emitting Diodes)
When VitaGuard® is switched on, all LEDs light up for a short time so
that you can see they work properly. During this time, the alarm LED
first lights up red and then yellow.
4.4.1 Alarm
LED
In the event of a higher-priority alarm, i.e. a
physiological alarm, the alarm LED flashes red.
In the event of a medium-priority alarm, i.e. a
technical alarm, the alarm LED flashes yellow.
Fig. 21 Alarm LED
4.4.2 Heart and respiration LEDs
The LED with the heart symbol flashes with every heartbeat of the
patient. In other words, this LED flashes as fast as the
heart beats.
The LED with the lungs symbol lights up with every
detected breath of the patient when the patient does
not move and is not moved. In other words, this LED
flashes as fast as the patient breathes.
Fig. 22 Heart and respiration LEDs
These two flashing green LEDs show you even in complete darkness
that monitoring is activated.
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...