82 Information for the doctor and qualified medical staff
Warn your pacemaker patients that the displayed heart rate may
possibly be affected by stimulating pulses. Point out to the caregiv-
ers that they must carefully observe pacemaker patients.
Do not operate VitaGuard® near MRI devices (magnetic resonance
imaging) or other systems that generate strong electromagnetic
fields. The electrode leads, for example, can heat up by induction,
causing burns under the electrodes and fire in the cables. The strong
magnetic fields generated by magnetic resonance image devices can
cause permanent damage to VitaGuard®.
9.2
Info
display
The
Info
display quickly presents the doctor with a summary of the
monitor settings and data. Other
Info
windows can be accessed with
the direction keys
U
and
V
.
The current page number and the total number of pages are dis-
played in the top right. For example, “1/10” means “the first of ten
pages”.
9.2.1
Info\ Last status messages
The last status messages provide
information on the directly
preceding monitoring period.
Here you can see when and why
a message appeared.
Fig. 46 Info\ Last status messages