92 Information for the doctor and qualified medical staff
Trend
storage (automatically for max 72 hours)
Interval
storage (set in the
System
menu)
Long term
storage (automatically for max eight hours)
Protocol
storage (automatic)
The VitaGuard® display lets you view stored
Events
and stored
Trends
.
Long term
and
Protocol
storage can be evaluated on a PC
only. The “Summary of stored signals and data” on page 97 presents
the signals and sample rates for the respective memory. The currently
utilized memory capacity is shown on the
Info
display. The installed
memory can store up to 200 events of two minutes each.
9.5
Event
storage
Please bear in mind that you must wait for the
Post-alarm time
to
expire after an alarm has ended before the current alarm event can
be completely stored.
The time of occurrence and the
length of the stored data are
stored for every alarm event. The
section “Summary of stored
Events
” on page 95 explains the
symbols in the columns for each
physiological parameter
[
©
ª
(
©
)
(
ª
)
X
P
]
.
Fig. 60 List of the stored events
M/I
marks the episodes that have been stored via
Manual storage
or
Interval recording
set in the
System
menu. The minimum and
maximum values of the respective physiological parameters are seen
at the bottom of the window.
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......