Information for the doctor and qualified medical staff 79
9 Information for the doctor and
qualified medical staff
The treating doctor is responsible for monitoring with VitaGuard®.
This also applies to ambulatory monitoring.
This section contains all safety and settings information that only
the treating doctor can make decisions on. Remember that all the
information and instructions in the sections “Intended use” on page
14 and “Safety” on page 20 must also be observed.
Only getemed AG personnel or authorized dealers certified by
getemed AG as medical product advisers in accordance with § 31
MPG (German Medical Products Act) may instruct the doctor and the
qualified medical staff on how to handle and use VitaGuard®. This
certification is awarded only to those persons that have received
adequate training from getemed AG for its products.
9.1 Safety
instructions
The safety instructions in this section address special technical and
medical issues that are of particular importance to the doctor and
qualified medical staff.
9.1.1 Preparing for a new patient
When more than one VitaGuard® monitor with differing settings are
used in the same environment, there is a risk of mixing monitors
and a particular patient may be monitored with unsuitable settings.
For this reason check the currently set alarm limits every time the
monitor is switched on.
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......