62 Alarms, displays, and views during monitoring
7.9
Heart rate and apnea monitoring
After the monitor is switched on, it may take up to twenty seconds
before the first values are displayed.
7.9.1 Heart rate alarms
The currently set alarm limits are always displayed.
When the displayed
Heart
rate
falls below the
Lower limit
for longer than the set
Bradycardia
delay
or
exceeds the
Upper limit
for longer than the set
Tachycardia
delay
or when the ECG signal is not detected for longer than the set
Asystole delay
VitaGuard® emits an acoustic alarm signal and displays the corre-
sponding message. The violated alarm limit and the alarm LED flash.
Go immediately to the patient when an alarm is reported and check
the patient’s condition.
The alarm is ended automatically when the heart rate returns within
the permitted limits.
Deviation alarms can also be activated in addition to the alarms
based on permanently set limits.
7.9.2 Apnea
alarms
An alarm is reported when apnea, i.e. respiratory arrest, is detected
for longer than the set
Apnea delay
. An alarm message then appears
on the display, the alarm LED flashes, and an acoustic warning is
emitted.