74 Alarm and monitor settings
8.5.7
System\ Settings protection
On,
Limited, Off
The codes that protect the alarm defaults from unauthorized
changes must be given by the doctor to those persons only whom
the doctor judges to be adequately informed about monitoring and
their responsibility for the patient. The doctor should point out that
the code must be treated as confidential, that settings should be
changed at the doctor’s request only, and that all changes must be
confirmed by the doctor.
VitaGuard® provides the following three settings for
Settings pro-
tection.
Settings protection ON
deactivates all options to change moni-
tor settings. The display presents only
View
1, the
Info
display,
and the
System
menu.
Settings protection Limited
enables access to all views and
menus. Of all the monitor settings, however, only the alarm limits
can be changed.
Settings protection Off
enables all views and menus and allows
changes to all monitor settings.
The factory setting is
Settings protection Limited
.
After highlighting the function
Settings protection,
press the <En-
ter> key to open a submenu.
This submenu always displays
Settings protection
as “00” irre-
spectively of the current setting. Pressing the <Enter> key acti-
vates
Settings protection
.
When you enter a code
Settings protection
appears as
Limited.
When you enter a different code,
Settings protection
appears as
Off
.
When the wrong code has been entered three times
Settings pro-
tection
cannot be so easily deactivated. In this case, consult your
authorized dealer.
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......