100 Information for the doctor and qualified medical staff
III BK-YE, 3 (black–yellow, 3 electrodes)
I YE-RD, 2 (yellow–red, 2 electrodes)
Here you can determine which lead is
used with two or three electrodes for
detecting the ECG signal (explanations
in the following).
Pacer detection ....................
Activate this for pacemaker patients so
that stimulating pacemaker pulses are
not mistakenly processed as R waves;
explanations can be found under
“VitaGuard® and other medical devices”
on page 81.
Heart rate alarms ...............
– When
Heart rate alarms
is set to
Limits only
, acoustic alarms are re-
ported when the measured values vio-
late the set alarm limits.
– When
Heart rate alarms
is set to
Limits & trends
, acoustic alarms are re-
ported when the measured values vio-
late the set alarm limits AND when
there is a positive or negative deviation
from the average heart rate measured
over the set interval.