104 Information for the doctor and qualified medical staff
Period
T2 (
Resp.
) .................... 4, 6, 8 ...
20
... 28, 30 seconds
When resumed respiration lasts longer
than T2, periodicity is no longer assumed.
Number of periods
.............. 2,
3
... 5, 6 periods
When the set number of periodic cycles
has been reached, the data are stored as
a silent alarm.
Min. respiration rate
......... 4, 5, 6 ... 9,
10
/min
VitaGuard® discards signals with a
lower rate of occurrence.
NOTE Changing the min respiration rate to 5/min has an effect only
when the age group has been set to
0 to 2 years
under
System\ Ad-
mit new patient
(see “
System\ Admit new patient
– restoring factory
settings” on page 88). For the age group
2 to 6 years
the min respir-
ation rate is fixed at 5/min, for the age group >
6 years
at 4/min.