42 Description
4.3.1 Direction
keys
With the direction keys you navigate from one
window to the next.
The direction keys also allow you to navigate within
the menu structure.
Fig. 18 Direction keys
4.3.2 <Enter>
key
The <Enter> key switches VitaGuard® on and off.
The <Enter> key also lets you confirm changes to
the monitor settings.
Fig. 19 <Enter> key
4.3.3 <Esc>
key
When an alarm is triggered, the <Esc> key serves to deactivate the
acoustic alarm signal for a set alarm mute time. During an alarm
condition the red alarm LED and the violated alarm limit flash. The
acoustic alarm is again emitted if the alarm condition persists af-
ter the alarm mute time has expired. Pressing the
<Esc> key during the alarm mute time a second
time reactivates the acoustic alarm.
Fig. 20 <Esc> key
Also when an alarm has automatically ended (because the vital
functions have restabilized by themselves) the alarm LED and the
violated alarm limit continue to flash until you press the <Esc>
key. The alarm LED, however, flashes slower than during an alarm.