90 Information for the doctor and qualified medical staff
The table lists the factory settings for each age group:
0
to
2
years
2
to
6
years
>
6
years
Min. respiration rate [/min]
10 5 4
Lower heart rate limit [/min]
80
60 55
Upper heart rate limit [/min]
220
150 120
All other settings are not specific to age groups.
The values set last are retained when VitaGuard® is switched off and
when the supply network or batteries fail to provide power.
9.3.4
System\ Pre-
and
Post-alarm time
In the event of an alarm, data for the pre-alarm and post-alarm times
set here are stored in addition to the duration of the alarm. These
times can be set from 30 to 180 seconds in steps of ten seconds.
Also the menu item “
System\ Interval recording”
explained on page 91
records data for the duration of the pre-alarm and post-alarm times.
9.3.5
System\ Alarm mute time
In order to allow caregivers to tend to a patient in peace during an
alarm, the acoustic alarm signal can be temporarily deactivated with
the <Esc> key. Set the
Alarm mute time
to 30, 60, 90, or 120 seconds
as required. The acoustic alarm signal is automatically reactivated
after this time.
9.3.6
System\ Date/ time
How to change these values is explained in the section “Changing
multiple-component settings” on page 87. This is used to set the
current date and the current time, e.g. for summer and winter times.
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......