46 Steps before and after monitoring
5 Steps before and after monitoring
The following summary shows you all the necessary measures that
need to be taken before monitoring. Also read information on how
VitaGuard® is switched on and off.
The doctor and the qualified medical staff are responsible for all
other important activities when “Preparing for a new patient” (see
page 79).
5.1
Summary of steps before monitoring
Insert the battery or batteries (do not switch on yet!).
Use the external power adapter to connect VitaGuard® to the
supply network (do not switch on yet!).
Connect the ECG patient cable to VitaGuard®.
Attach the ECG electrodes to the patient.
Connect the ECG electrodes to the patient cable.
Switch on VitaGuard® as explained in the next section.
Make sure that after the monitor is switched on the indicator
lamps light up briefly and a short sound is emitted by the alarm
buzzers.
Check that the alarm limits displayed are the same as those
recommended by the doctor.