Steps before and after monitoring 47
5.2 Switching
on
Press the <Enter> key for several seconds to switch on VitaGuard®.
In the first minute of operation no acoustic signals are emitted so
that you have time to check all cables. The alarm bell is crossed out
for this time and the remaining time is shown next to it. Text mes-
sages, on the other hand, are shown from the beginning.
When no patient cable is connected, an acoustic reminder signal is
emitted as a short tone
every twenty seconds after the monitor is
switched on. The technical alarms for cable and electrode monitoring
are not activated until the patient cable is connected and the first
plausible data have been calculated. A text message in the status line
reports from the beginning that the cables are being checked.
After the device has been switched on, the following displays and
signals show you that the monitoring system is fully operable.
All indicator LEDs light up briefly. During this time the alarm LED
first lights up red and then yellow.
A brief tone is emitted to indicate that the acoustic alarm buzzer
is fully operable.
If the alarm buzzer does not emit the acoustic signal after the device
has been switched on, you must immediately send VitaGuard® to
getemed AG or your authorized dealer for inspection. Please consult
your authorized dealer for a replacement device.
Observe the patient carefully until the replacement device arrives.
Bear in mind that the patient is not being monitored at this time
and that no alarm will be reported in an emergency.
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......