98 Information for the doctor and qualified medical staff
9.10 Settings in the
Heart rate
menu
(Settings
protection Off)
For this setting, Settings protec-
tion must be set to Off as ex-
plained under “System\ Settings
protection On, Limited, Off” on
page 73.
The possible settings when
Settings protection
is set to
Limited
are explained in the
section “
Heart rate
menu –
alarm settings (Settings
protec-
tion Limited)
” on page 75.
Factory settings are shown in
bold type.
Fig. 68 Settings in the Heart rate menu
Silent lower limit (Heart)
30
, 35 ... 50 ... 175, 180/min
Lower limit for the heart rate; when the
measured value falls below this limit for
longer than the set
Bradycardia delay,
a silent alarm is stored.
Silent upper limit (Heart)
. 100, 105 ...
255
/min
See
Silent lower limit.
Bradycardia delay
.............. 1, 2, 3, 4, 5,
6
... 14, 15 seconds
Delay between when bradycardia is de-
tected and the corresponding alarm is
triggered.
Tachycardia delay
............... 1, 2, 3, 4, 5 ...
15
... 23, 24 seconds
See
Bradycardia delay
, but for tachy-
cardia.