30 Safety
3.7
Safety with proper maintenance only
VitaGuard® can operate safely and reliably over the long term only
when it is subject to proper maintenance and use.
Check visually for any damage on VitaGuard®, the patient cables
including the connections, the external power adapter, and the
electrodes every time you use VitaGuard® for monitoring.
Every eighteen months at the latest VitaGuard® and accessories
must be serviced by getemed AG to comply with safety regulations.
Repairs must be performed by getemed
AG only. Clarify the neces-
sary procedure with your authorized dealer.
For the protection of our service personnel, disinfect VitaGuard® and
the patient cables with Virkon®, available as a spray or wiping solu-
tion, before sending them to getemed AG.
3.7.1 Cleaning VitaGuard® and accessories
Before cleaning VitaGuard®, remove the batteries.
Before cleaning VitaGuard®, detach the electrodes and cables from
the monitor and from the patient.
Do not under any circumstances use solvents like ether, acetone, or
benzene. These substances can cause malfunctions and attack the
housing plastic.
Also, do not use any cleaning agents containing abrasive substances
and no coarse brushes or hard objects.
VitaGuard® and accessories can be cleaned any number of times
when the recommended cleaning agents are used.
VitaGuard® and accessories must not be sterilized.
VitaGuard® and the cable plugs must not be immersed or otherwise
penetrated by liquid.
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......