88 Information for the doctor and qualified medical staff
After changing a value, pressing the <Enter> key a second time causes
a prompt to appear with
Accept: No
highlighted. Press the
Y
key to
highlight
Accept: Yes
. Confirming the prompt
Accept: Yes
with the
<Enter> key displays the changed value in the list. To exit the menu
press the <Esc> key.
9.3.2
System
\
Operating area: Home
or
Clinic
Operating area
lets you decide whether the value you have entered
for
Settings protection
is retained the next time VitaGuard® is
switched on.
When you select
Home
,
Settings protection
is
On
when the device is
next switched on.
When you select
Clinic
,
Settings
protection
is set to the selected
value when the device is next
switched on.
Fig. 55 Operating area: Home or Clinic
9.3.3
System\ Admit new patient
– restoring factory
settings
IMPORTANT: This deletes all stored data and all monitor settings for
a specific patient.
Check that the new monitor settings are suitable for the patient.
All data are deleted and all user settings are restored to the factory
values, so you are prompted
whether you want to
Continue
.
Press the
Y
key to highlight
Accept:
Yes
and then the <Enter> key.
Fig. 56 Warning before changes under System\ Admit new patient
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......