80 Information for the doctor and qualified medical staff
It is important that VitaGuard® is configured so that false alarms are
avoided to the greatest possible extent. Frequent false alarms can
prove detrimental to the alertness of caregivers.
When VitaGuard® is to be used for a new patient, the doctor or the
qualified medical staff are obliged to take the following important
precautionary measures.
Place used consumables such as electrodes
in a plastic bag before
disposing of them in household or medical waste.
Clean the device and disinfect all cables (e.g. as described in the
guidelines from the Robert Koch Institute).
Insert new batteries or a fully charged block battery.
Select the age group in
System\ Admit new patient
as explained
under “
System\ Admit new patient
– restoring factory settings”
on page 88.
Check that the monitor settings are suitable for the patient and, if
necessary, adapt them.
Consider that the monitor settings may need to be changed at a
future date and, as appropriate, arrange appointments to change
these settings.
Check that the acoustic alarm signal is loud enough to be heard
over the prevailing or expected noise levels in the monitor’s envi-
ronment.
When necessary, set
Settings protection
to
Limited
.
When necessary, train caregivers in the necessary resuscitation
measures.
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......