Information for the doctor and qualified medical staff 103
9.12 Settings in the
Respiration
menu
(Settings
protection Off)
For these settings,
Settings protection
must be set to
Off
as ex-
plained under “
System\ Settings protection On, Limited, Off
” on
page 74. The possible set-
tings when
Settings protec-
tion
is set to
Limited
are
explained in the section
“
Respiration
menu – alarm
settings
(Settings protecti-
on Limited)
” on page 78.
Factory settings are shown in
bold type.
Fig. 70 Settings in the Respiration menu
Silent apnea delay
.............. 8, 10 ... 22,
34
seconds
Same as
Apnea delay
, but a silent alarm
is stored when exceeded.
Periodic respiration
.......... Yes/
No
This activates the detection of periodic
respiration.
NOTE Activating the monitor for periodic respiration only operates if
the age group 0–2 years has been previously selected.
Fig. 71 Periodic respiration
Period
T1 (
Delay
) .................
6
, 8, 10 ... 18, 20 seconds
For the detection of periodic respiration
apnea must be longer than T1, but
shorter than the set
Apnea delay
.