66 Alarms, displays, and views during monitoring
Message
Meaning
Cause or elimination
No.
Conflicting
HR limits!!
The heart rate’s
Lower
limit
has been set higher
than the
Upper limit.
-
Correct the heart rate limits.
19
Corrupted
ECG signal!!
The ECG signal is
corrupted too strongly by
50 Hz interference
signals from the supply
network.
-
Attach the electrodes as symmetrically
as possible.
-
Replace the electrodes.
-
Select the ”I YE-RD, 3” lead.
-
Proceed in accordance with the section
“Electromagnetic interference” on page 24.
24
Hardware
fault!!
The monitor has detected
an internal fault.
-
Switch off the monitor, wait for thirty
seconds, and switch it back on: if this
message persists, the monitor is defect.
14
Internal
data error!!
The internal software
monitor has detected a
data transfer error.
-
Switch off the monitor, wait for thirty
seconds, and switch it back on: if this
message persists, the monitor is defect.
18
No cables
connected!!
The monitor discovers
that the patient cable is
not connected.
-
Connect the patient cable.
21
No power
adapter !!
The power adapter has
been disconnected.
-
Reconnect the external power adapter
or press the <Esc> key.
17
Recharge
battery!!
The battery voltage is too
low: the monitor can no
longer operate reliably
-
Operate the monitor with the external
power adapter to recharge the block
battery, or insert non-rechargeable
batteries.
32
Replace
batteries!!
The battery voltage is too
low: the monitor can no
longer operate reliably.
-
Insert new batteries or a new block
battery or operate the monitor with
the external power adapter.
33
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......