84 Information for the doctor and qualified medical staff
9.2.3
Info\ Measurements: HR & Resp.
Info\ Measurements: HR & resp.
displays the various average
heart rate values calculated since
the monitor was switched on.
These values are lost when the
monitor is switched off.
Fig. 48 Info\ Measurements: HR & Resp.
HR: Average
is calculated over of the
Averaging interval
set in the
Heart rate
menu.
HR: Current deviation
shows how much the current heart rate
deviates from the
Average
in percent.
This
deviation
is used for triggering deviation alarms when
Heart
rate alarms
has been set to
Limits & trends
in the
Heart rate
menu.
This setting is restored as the factory setting when the
Admit new
patient
function is executed in the
System
menu.
Periodic respiration
displays the time in percent that
Periodic
respiration
has been detected since the monitor was switched on.
This value is displayed only when the function
Periodic respiration
is
set to
On
in the
Respiration
menu and when
Age group
is set to
0 to
2 years
under
System\ Admit new Patient
.
Basal impedance
displays the value measured between the yellow and
the red electrode (see “Checking the
basal impedance
” on page 55).
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......