Information for the doctor and qualified medical staff 91
9.3.7
System\ Language
The menu option
Language
is marked with a flag symbol
in the event that you do not understand the set language.
The submenu
Language
lets you choose between the languages
supported by your monitor version.
9.3.8
System\ Analog input
1 + 2
You can activate and deactivate the two analog inputs separately.
Both analog inputs have an input range from 0 to 2.5 V. An analog
signal at input 1 (when activated) is scanned with 1 Hz and stored, a
signal at input 2 is scanned with 32 Hz and stored.
9.3.9
System\ Interval recording
The doctor can use
Interval recording
for specific situations and can
set an interval in steps of ten minutes after which monitored data are
again stored. After each of these intervals, the monitor stores data for
the
Pre-alarm
and
Post-alarm times
. The setting
0 min
deactivates
interval recording.
9.4 Data
storage
functions
The function
Admit new Patient
in the
System
menu overwrites all
the currently stored data and restores the factory settings. If neces-
sary, transfer the data beforehand to a PC.
The memory contents of VitaGuard® are also retained when the
power adapter or batteries fail.
VitaGuard® features the following data storage functions:
Event
storage (automatic storage of
Alarms
and
Silent Alarms
or
Manual storage)
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......