Preparing for heart rate and apnea monitoring 49
6 Preparing for heart rate and apnea
monitoring
This section is divided into the following parts.
Safety information when monitoring heart rate and apnea
Connecting electrodes, the patient cable, and VitaGuard®
Technical alarm from the electrode contact monitor
Determining the optimal electrode configuration
ECG lead, electrode color coding
Optimizing the heart and respiration signals –
signal amplitudes in
View
1
Checking the
basal impedance
6.1
Safety information when monitoring heart
rate and apnea
Observe the following points before monitoring with VitaGuard®.
CAUTION Interference signals can prevent a heart rate alarm from
being reported when under certain unfavorable conditions the
monitor misinterprets these interference signals as heart signals.
Interference signals can originate from the power supply or electri-
cal apparatus in the monitor’s environment. Observe the instruc-
tions under “Electromagnetic interference” on page 24.
The doctor can deactivate the
Apnea alarms
as described in the
section “Settings in the
Respiration
menu
(Settings
protection
Off)
” on page 103. The
Apnea alarms
are then no longer activated
and
Off
is displayed instead of the respiration rate. When a new age
Содержание VitaGuard VG2100
Страница 1: ...VitaGuard VG2100 Apnea and Heart Rate Monitor Operating instructions...
Страница 2: ......
Страница 4: ......
Страница 10: ...Table of contents...
Страница 111: ...Evaluating stored data on a PC 111 Fig 72 VitaWin register Events in graph form...
Страница 122: ...122 Table of figures...
Страница 123: ......