78 Alarm and monitor settings
8.7.2
Respiration
menu – alarm settings (
Settings
protection Limited
)
These settings can be changed
only when
Settings protection
is set to
Limited
in the
System
menu.
Factory settings are shown in
bold type.
Fig. 45 Respiration menu for viewing and setting alarm limits
Apnea delay
............................ 8, 10 ...
20
... 32, 34 seconds
VitaGuard® interprets apnea and trig-
gers an alarm when a respiration signal
or movement is not detected and
Apnea
delay
is exceeded. The doctor must
have set
Apnea alarms
to
Always
.
Apnea alarms
........................
Off
/
Always
CAUTION A deactivated apnea monitor can no longer detect apnea!
O
ff
and
Always
switches off and on
the apnea monitor. In both cases the
respiration waveform is displayed and
stored.